Essentials in Bioassay Design and Relative Potency Determination

Publication
Article
BioPharm InternationalBioPharm International-04-01-2016
Volume 29
Issue 4
Pages: 49–52

The author describes common components of a relative potency bioassay and provides a framework for assay development, calculation, and control.

In defining bioassays and relative potency, the United States Pharmacopeia (USP) states (1):

“Because of the inherent variability in biological test systems (including that from animals, cells, instruments, reagents, and day-to-day and between-lab), an absolute measure of potency is more variable than a measure of activity relative to a Standard. This has led to the adoption of the relative potency methodology. Assuming that the Standard and Test materials are biologically similar, statistical similarity (a consequence of the Test and Standard similarity) should be present, and the Test sample can be expected to behave like a concentration or dilution of the Standard. Relative potency is a unitless measure obtained from a comparison of the dose-response relationships of Test and Standard drug preparations.”

The purpose of this article is to provide a presentation of the most common components of a relative potency bioassay and to provide a framework or platform for assay development, calculation, and control. It is not meant to be mutually exclusive or mutually exhaustive of all the issues associated with a bioassay, but it covers the essential components in relative potency calculation and reporting.

Bioassay purpose and fitness for use
Bioassays can be in vitro, in vivo, or ex vivo and can be used for a variety of purposes. Those purposes include the following:

  • Process development, process characterization, and product development

  • Product release testing for drug substance or drug product

  • In-process control and intermediates testing

  • Stability and product integrity testing.

Bioassay dose response
Getting the bioassay dose response dilution scheme correct is one of the most important aspects of a bioassay. There are many dose response strategies that may be used, such as the following:

  • Single dose (not recommended)

  • Two dose (not recommended)

  • Three dose (not recommended)

  • Four dose (good but difficult to mask for saturated doses)

  • Five dose (good for parallel line analysis [PLA] and dose masking)

  • Fine dose (see Figure 1) or more (good for four parameter [4PL] fitting and dose masking).

Figure 1: Nine-dose sigmoidal curve. [courtesy of author]

Generally, one-, two-, and three-dose schemes should be avoided as they are difficult to control variation and will be problematic during relative potency calculation. The ideal dose response curve is a five-point curve for parallel line analysis (linear) and a nine-point dose scheme or more for sigmoidal fits. For a sigmoidal curve, ideally it should be three points in the inactive or low response section of the curve, three points in the linear part of the curve and three points in the saturated area of the curve on the log or log 10 scale from a serial dilution. In a parallel line analysis (non-sigmoidal), a five-dose scheme is ideal in the linear part of the curve.

4PL versus 5PL fitting
If fitting a logistics sigmoidal curve, there are two common curve-fitting techniques: four parameter (4PL) and five parameter (5PL). Four-parameter curves assume symmetry in the upper and lower region of the curve fit, and five-parameter curves assume the curve shapes of the upper and lower asymptote are different. An evaluation of the residuals from the 4PL fit will indicate if the curve fitting technique is appropriate or if a 5PL fit is warranted. Residuals should be random and normal and show no systematic pattern across the dose response if the 4PL fit is good.

Outlier analysis
In discussing outliers (see Figure 2), USP <1032> states, “The statistical elements of bioassay development include the type of data, the measure of response at varying concentration, the assay design, the statistical model, pre-analysis treatment of the data, methods of data analysis, suitability testing, and outlier analysis. These form the constituents of the bioassay system that will be used to estimate the potency of a Test sample.”

Figure 2: Nine-dose 4PL with outlier. [courtesy of author]



The following are recommended approaches for outlier detection and removal (2):

  • Studentized residuals from the curve fit is greater than 2.575 (99% risk)

  • z score of the removed point of 3 or more (99.73% risk).


The influence of the removed point on the curve can be evaluated by examination of the change in R2 and root mean squared error (RMSE) of the residual error. Generally, no more than one point or one dose on the dose curve should be removed.

 

 

Transformation and curve weighting
Data transformation may be done in both the dilution and the measured response if it can be shown to improve the linearity of the assay. USP <1034> (3) states, “Fit the statistical model for detection of potential outliers, as chosen during development, including any weighting and transformation. This is done first without assuming similarity of the Test and Standard curves but should include important elements of the design structure, ideally using a model that makes fewer assumptions about the functional form of the response than the model used to assess similarity.”

Weighting reduces the variation in relative potency determination if the variation along the dose response lacks homogeneity. Typically, 1/variance of the residuals at each dose is used when the variation is statistically different. An unequal variances test using Bartlett’s or Levene’s can be used to determine if the variances are statistically different. Fixed 1/variance weighting is generally the recommended approach. Weighting is applied to both reference and test.

Masking for parallel line analysis
Masking or removal of the saturated measurement (see Figure 3) in a linear parallel line analysis is recommended. Only the high or low (saturated) dose of the response curve is removed, and sequential three or four doses are used for relative potency determination depending on the dosing scheme. Masking will remove the saturation (hockey stick effect) in the parallel line analysis.

Figure 3: Hockey sticking (saturated) dose response for a linear fit. [courtesy of author]

Systems suitability and evaluating parallelism
There are several techniques for the evaluation of parallelism. Table I provides a summary of methods for the evaluation of parallelism. Other systems suitability criteria include the measurement of a known control, relative variation (%CV) of repeated measures, and curve depth for signal strength.

Table I: Evaluation of parallelism and systems suitability.

Table I: Evaluation of parallelism and systems suitability.

Calculation of relative potency
The calculation of relative potency is shown in Equation 1 and may change depending on the assay.

EC50 is calculated from the 4PL constrained inflection point and back transformed (antilog) using the following equations depending if it is log 10 or log transformed in the dilution factor (see Equation 2).

Controlling the influence of the standard in relative potency
Notice in Equation 1 that there is a correction factor (CF) applied to the relative potency determination. CF reference is a correction factor for qualification of a new reference to correct for shifts in potency due to a change in reference. CF reference stability is a correction for the loss of potency of the reference over time. Failure to include these correction factors in the potency calculator will cause the assay to become hyper or hypo potent overtime.

Work flow in the potency assayFigure 4 presents a good workflow for a bioassay. Data from the bioassay should be quality control checked, and all reportable results should be sent to a laboratory information management system (LIMS). Key parameters and reagents from the bioassay should be used for tracking and trending.

Figure 4: Bioassay workflow. LIMS is laboratory information management system. [courtesy of the author]

Monitoring and controlling relative potency
It is critical to monitor the performance of the bioassay over time. The unconstrained EC50 standard is generally considered the best measure of the stability and consistency of the assay. To detect if assay drift is a method issue or standard issue a comparison to the unconstrained EC50 should be made. If both the reference and the test are drifting, it is change in the assay over time but no impact to relative potency. If the test is stable and the reference is trending it will impact relative potency and should be corrected or a new reference selected.

Conclusion
Bioassays are powerful tests of relative potency and biological activity but need care in the design of the assay and the analysis to make sure they are a reliable indication of the change in potency. Validation of the assay (4) will demonstrate the accuracy, repeatability, and linearity of the assay and its fit for use in the measurement of relative potency. Care in the selection of the dose response, outlier detection and removal, masking, transformation, and weighting will make the assay more stable and repeatable.

References
1. USP, <1032> Design and Development of Biological Assay, USP 38–NF 33 (USP, December 2015).
2. T. Little, BioPharm Intern 29 (1) (January 2016).
3. USP, <1034> Analysis of Biological Assays, USP 38–NF 33 (USP, December 2015).
4. USP, <1033> Biological Assay Validation, USP 38–NF 33 (USP, December 2015).

Article DetailsBioPharm International
Vol. 29, No. 4
Pages: 49–52

Citation: When referring to this article, please refer to it as T. Little, "Essentials in Bioassay Design and Relative Potency Determination," BioPharm International 29 (4) 2016.

 

Recent Videos
Lee Cronin, founder and CEO of Chemify
Jay Rajagopalan, senior director—Engineering & Product Management for Malema at PSG Biotech | Image Credit: BioPharm International
© 2024 MJH Life Sciences

All rights reserved.